MODEL 14C SURVEY METER WITH END-WINDOW GM PROBE
Satisfies NRC requirements for Nuclear Medicine Departments
- Monitors alpha, beta, and gamma
- 0-2 R/hr range
- External end window GM probe (39" cable included)
- Internal energy-compensated GM detector
- Five counting scales (x0.1, x1, x10, x100, x1000)
- Alpha Efficiency – 15%• Beta Efficiency – 5% for C-14, 20% for Sr-90
- Gamma Efficiency – 14% for Tc-99
- Sensitivity – 2100 cpm/mR/hr for Cs-137
A standard survey instrument for nuclear medicine departments, the portable Model 14C Survey Meter with End-window Probe meets the essential surveying needs of most Nuclear Medicine facilities. The end-window probe is used to check hands, clothing, packages, and other surfaces for contamination, and for background readings. For high-range gamma detection, the built-in energy-compensated GM detector with the x1000 multiplier range handles counts up to 2R/hr. The meter features a built-in speaker with ON/OFFswitch, front-access calibration controls, push-button reset to zero the meter, and a toggle switch to select FAST (4 sec.) or SLOW (22 sec.) response.
SPECIFICATIONS
Survey Meter
- Compatible Detectors: GM, Scintillation
- Meter Face: 0-2 mR/hr, 0-4.2k cpm, battery test
- Threshold: 30 mV ± 10 mV
- Multiplier Ranges: x0.1; x1; x10; x100 for external detector; x1000 for internal detector
- Internal Detector: Energy-compensated GM, for high range gamma detection only; 2000 mR/hr
- High Voltage: 900V
- Response: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) for 90% of final reading
- Reset: Push button to zero meter after over-range exposure
- Audio: Built-in unimorph speaker with ON/OFF switch
- Sensitivity: 2100 cpm/mR/hr for Cs-137
- Batteries: 2 each, size “D”, typical life 600 hours
- Construction: Painted aluminum housing
- Dimensions: 3.5" w x 6.5" h x 8.5" l (9 x 16.5 x 22 cm)
- Weight: 3.5 lb (1.6 kg) including batteries
End-Window GM Probe
- Indicated Use: Alpha, beta, and gamma surveying; sample counting
- Detector: End-window halogen quenched GM
- Window: 1.7 ±0.3 mg/cm² mica
- Window Area: 6 cm² active, 5 cm² open
- Efficiency (4pi): 2% for C-14, 10% for Sr-90 and Y-90, 7% for Tc-99, 7% for Pu-239
- Gamma Sensitivity: 2100 cpm/mR/hr for Cs-137
- Energy Response: Energy dependent
- Operating Voltage: 900 volts
- Protective Screen: 79% open, stainless steel
- Housing: Anodized aluminum
- Dimensions: 1.8" diameter x 5.8" l (4.6 x 14.7 cm)
- Weight: 1 lb (0.5 kg)
- Detector Cable: 39" (99 cm), type “C” connectors
051-014 - Survey Meter, Model 14C Includes Internal Energy-Compensated GM Detector, External End-Window GM Probe, Cable and mounted Check Source (Cs-137, 1.0 µCi).