Pro-NM Performance SMALL

[ click the picture ]

Nuclemed - click to enlarge

Phantom for small FOV NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). Among other things, it can be used to evaluate: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution,  construction filters on spatial resolution, attenuation and scatter compensation.

Technical data (can be modified to customer specifications):

  • main cylinder:
    – inside cylinder diameter: 140mm
    – inside cylinder height: 150mm
    – cylinder wall thickness: 6mm
  • cold rods insert:
    –rod diameters: 3.2, 4.8, 6.4, 7.9, 9.5 and 11.1mm – height of rods: 40 mm
  • cold spheres:
    - solid sphere diameters: 3.2, 4.8, 9.5, 12.7, 15.9 and 19.1mm – height of the center of the spheres from the base plate: 78.5 mm
  • optional PET Lid with cylindrical samples:
    - refillable thin-walled cylinders, diameters: 4, 6, 8, 12, 16mm
    - PTFE solid cylinder diameter: 16 mm
    - cylinder height: 25.4 mm
  • optional heavy duty carrying case

Product features:

  • Complies with:
    – NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    – AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    – AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    – ACR–SNM  (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration

 

08-111 - Pro-NM Performance SMALL - standard version
08-112 - Pro-NM Performance SMALL - version with the PET Lid