Pro-NM Performance SMALL
[ click the picture ]
Phantom for small FOV NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). Among other things, it can be used to evaluate: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, construction filters on spatial resolution, attenuation and scatter compensation.
Technical data (can be modified to customer specifications):
- main cylinder:
– inside cylinder diameter: 140mm
– inside cylinder height: 150mm
– cylinder wall thickness: 6mm - cold rods insert:
–rod diameters: 3.2, 4.8, 6.4, 7.9, 9.5 and 11.1mm – height of rods: 40 mm - cold spheres:
- solid sphere diameters: 3.2, 4.8, 9.5, 12.7, 15.9 and 19.1mm – height of the center of the spheres from the base plate: 78.5 mm - optional PET Lid with cylindrical samples:
- refillable thin-walled cylinders, diameters: 4, 6, 8, 12, 16mm
- PTFE solid cylinder diameter: 16 mm
- cylinder height: 25.4 mm - optional heavy duty carrying case
Product features:
- Complies with:
– NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
– AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
– AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
– ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals - CE certified
- the Manual provides detailed guidelines for carrying out each test, results assessment and registration
08-111 - Pro-NM Performance SMALL - standard version
08-112 - Pro-NM Performance SMALL - version with the PET Lid