Radhound - Alarm
A low cost digital area monitor available with a range of probe options for all your contamination monitoring and radiation protection requirements. The Radhound Alarm has a large digital display, the option for mains or battery operation and can be easily wall mounted. The easy to use password protected configuration menu enables a choice of probe and alarm thresholds to be setup.
Alarm output is provided by remotely mounted audio sounder and visual beacon.
Low cost.
Simple to operate.
Settable alarm.
Audible count rate (with on/off switch).
SPECIFICATIONS
Units CPS, CPM, µSv/hr with autorange
Display Clear backlit LCD display
Controls Power, up, down, OK (menu keys)
Environmental
Operating temperature -10°C to +50°C
Storage temperature -25°C to +60°C
In conformity with EMC directive (89/336/EEC) as amended by Directive 92/31/EEC - Low Voltage Directive (73/23/EEC), EN61326-1,
EN61000-3-2, EN6100-3-3. (CE mark). Designed to meet IEC 60325-2006 and IEC 60846-2004 Nuclear Instrumentation Standards
Power
Lithium Ion rechargeable, typically >12 hours continuous use.
Charger supplied
Cleaning Radhound is chemical resistant, and can be cleaned with Alcohol wipes
Mechanical
Dimensions: (H x W x D) 250 x 250 x 150 mm
Max Approx (with stand)
Range Typically 0 to 99,999 counts per second (range and units are software selectable depending on probe)
Response
Time 0.5 seconds
Averaging: 5, 10, 15 seconds
Integration time up to 24 hrs
Functionality Rate, histogram, timed count, count to time, time remaining to dose, alpha beta discrimination (depending on probe)
High Voltage 350 - 1200V selectable in menu (can be locked)
Averaging ‘Smart Averaging’ provides fluid number change, whist retaining a response time adjustable between slow, medium and fast