Pro-Dose Small Water

[ click the picture ]

Nuclemed - click to enlarge Nuclemed - click to enlarge Nuclemed - click to enlarge

Stationary water phantom for high energy photon dosimetry with all types of ionization chambers.

Technical data (can be modified to customer specifications):

  • dimensions: 200 x 200 x 110 mm
  • cover made of PMMA
  • solid construction ensures durability and no leaks
  • entrance wall made of milky / opaque 3mm thick PMMA
  • fixed measuring depth at 50mm
  • positioning markers on the top and side surfaces
  • phantom can be filled with water through a seal plug - a funnel is provided to facilitate the process
  • two elastic extension vessels compensate changes of the water volume caused by ambient temperature changes
  • well fitted holders are available for different chamber types
  • chamber holders can be interchanged (option)
  • heavy duty carrying case

Product highlights:

  • opaque / white entrance wall with crosshair, makes it very easy to accurately position the phantom in the beam
  • side markers at the measurement depth provide further assistance during positioning
  • interchangeable holders for different ionization chambers
  • extremely well fitted holders are very thin at the reference point – 1mm thick PMMA
  • water-proof ion chamber holders are also available

Product features:

  • complies with:
    – IAEA Technical Reports Series No. 398
    – IAEA Technical Reports Series No. 430
    – IAEA Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems (IAEA-TECDOC-1540)
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration

 

Pro-Dose Small Water 07-101