Cobalt-57 Rigid Line BM83-40

[ click the picture ]

Nuclemed - click to enlarge
  • Used to test SPECT camera head rotation alignment
  • RadQual offers Co-57 line sources in range of activities and dimensions to meet the different requirements of various camera manufacturers
    • Standard activity is 1.5 mCi; maximum source activity is 25 mCi
    • overall length 12.5 inches
    • active length 12.0 inches
    • overall diameter 0.25 inches
    • active diameter 0.04 inches
  • All sources are double encapsulated stainless steel
  • Sources shipped with Technical Data sheet that includes Radiation Safety Recommendations and Leak Test Results
  • The output linearity meets a specification of ±5% coefficient of variation as measured for 5 mm segments versus the mean value for all segments
  • SS&D NR-1235-S-107-S

BM83-40  Cobalt-57 Rigid Line