Tissue Mimicking Ultrasound QC Phantom

[ click the picture ]

Nuclemed - click to enlarge

The Tissue Mimicking QC Phantom 411 LE is designed to meet the needs of the small ultrasound lab. Affordably priced, this phantom provides a standard of quality assurance required for routine testing of ultrasound scanning systems.

The 411 LE features

  • one mid-depth axial resolution target at 6 cm with pins spaced at 2, 1 and 0.5 mm.
  • single 6 mm diameter anechoic cyst at 5 cm
  • all pin targets are constructed of 0.374 mm nylon
  • This economical phantom provides all the required quality control indicators for testing your ultrasound system:
    distance accuracy
    image uniformity
    axial and lateral resolution
    cyst imaging capabilities
    depth of penetration
    dead zone measurement

Specifications

Tissue-Mimicking Material
Speed of Sound 1540 + 10m/s
Attenuation 0.5 +0.05 dB/cm/MHz

Targets
Anechoic Cysts
Speed of Sound 1540 + 10m/s
Attenuation 0.5 +0.01 dB/cm/MHz

Pin Targets
Diameter 0.374 mm (0.012 in)
Size Tolerance + 5%
Location Tolerance +1 mm (0.004 in)

Container
Walls ABS
Scanning surface Composite Film
Size 18 x 15 x 7 cm (7 x 6 x 2.8 in)
Weight 2.45 kg (5.5 lbs)

Options - Foam lined carrying case

0695-0137 Tissue Mimicking UltrasoundQC Phantom 411 LE