Pro-MAM Accreditation FF

[ click the picture ]

Nuclemed - click to enlarge

This Full Field phantom was designed to test the performance of a digital mammographic system by evaluating the system’s ability to image small structures similar to those found clinically: micro-calcifications, fibrous structures in ducts and tumor-like masses. It is similar to the Pro-MAM Accreditation phantom, the main difference being the size: the FF (Full Field) version is larger and covers the entire image detector, thus eliminating scatter.

Technical data

(can be modified to customer specifications):

    dimensions: 311.2 x 190.5 x 41.3 mm
  • simulates 42 mm compressed breast of average glandular/adipose composition (50%/50%)
  • wax insert: – nylon fibrils diameters: 0.89, 0.75, 0.61, 0.54, 0.40 and 0.30 mm – microcalcifications: 0.33, 0.28, 0.23, 0.20, 0.17 and 0.14 mm Al2O3 specks – tumor-like masses: 1.00, 0.75, 0.50, 0.38, 0.25 and 0.20 mm thick
  • comfortable carrying case (optional)

Product features:

  • Complies with:
    – IEC 61223-3-2
    – ACR Mammography Accreditation Program Requirements
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration

 

Pro-MAM Accreditation FF - 03-502