Phantom for performance testing of mammography X-ray units. It has 49 holes that generate subtle changes in contrast. It can be used to detect small variations in system performance.

Technical data (can be modified to customer specifications):

  • dimensions: 62.7 x 62.7 x 62.7 mm
  • hole depth (7 columns): 0.853, 0.533, 0.332, 0.208, 0.129, 0.080, 0.050 mm
  • typical contrast for mammographic energies: 6.60, 4.20, 2.60, 1.70, 1.00, 0.65, 0.41%
  • diameter of holes (7 rows): 4.292, 2.524, 1.485, 0.873, 0.513, 0.302, 0.177 mm
  • optimized for digital units
  • includes a rotating support plate
  • comfortable carrying case

Product features:

  • complies with:
    – IEC 61223-3-2
    – ”European Guidelines for Quality Assurance in Mammography Screening”
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration

 

Pro-MAM Contrast - 03-105