DINMAM

[ click the picture ]

Nuclemed - click to enlarge

In accordance with DIN 6868 Part 7 and 152

Functionality

  • 2 Resolution test patterns (8, 10, 13, 16 lp/mm)
  • 7 Low contrast objects with 5.5 mm diameter
  • depths 0.10 — 0.40 mm in 6 steps of 0.05 mm
  • Reference point for optical density measurement
  • Cavity for the positioning of a dose detector
  • 2 Rows of 5 steel balls, used for testing the limitation of the image at the chest wall side
  • 3 Contrast levels
  • full cut out
  • 3 mm deep cut out
  • 3.5 mm PMMA + 2.5 mm Teflon
  • Orientation help by contrasting lines
  • For easy setup the absorption plates are fitted with alignment pins

This phantom is your ideal companion for both acceptance and constancy testing.

Overall dimensions: 240 x 180 x 46 mm

Made of 4 parts; one 6 mm structure plate and three 20 mm absorption plates, thereby complying with both DIN 6868 part 7 and part 152, saving you the need for a second phantom.